BodyGuard® ColorVision®
Epidural Infusion Pump

Deliver Epidural Analgesia Using a Drug Library and Color Display

Deliver your own custom drug protocols for continuous or patient-controlled epidural analgesia (PCEA). The BodyGuard ColorVision epidural infusion pump is color-coded for safety and makes information easy to read on a large full-color screen.

Key Features

  • Customizable Drug Library. Build your own protocols for epidural infusion therapy.
  • Multiple Epidural Infusion Options. Deliver either continuous infusion or patient-controlled epidural analgesia (PCEA) bolus doses, or both.
  • IEC Neuraxial-Compliant. Complies with IEC 80369-6 for neuraxial patient connections.

Main Features

  • Auto Bolus Option. You can deliver programmed intermittent dosing.
  • Large Full-Color Screen. Makes epidural therapy parameters easy to read.
  • Pole-Mount or Portable. Use one pump for hospital, clinic, or patient’s home.
  • Small and Lightweight. Easy to carry and fits in tight spaces.
  • Battery or AC-Powered. Runs on a rechargeable battery or AC (mains) power.
  • Automatic Backup Battery. When installed in the pole-mount charger, the pump switches from AC power to battery backup whenever power is interrupted.

Safety Features

  • MediGuard™ protection. Ensures that drug protocols are appropriate to the patient’s weight.
  • Color-Coded. The pump is yellow-labeled to distinguish it from other infusion systems.
  • Triple Free-Flow Protection. BodyGuard® tubing sets include a clamp and backcheck/anti-free-flow valve. A spring-loaded plate pinches the tubing when the pump door is closed.
  • Double Air-in-Line Protection. The pump’s ultrasonic air sensor triggers an alarm if an air bubble exceeds 5 mL or if the total volume of air over 15 minutes exceeds your specified volume (adjustable 0.1 to 1.0 mL).
  • Maximum Flow Rate Limit. Set this limit to protect against possible large programming errors. The limit applies to all infusion modes.
  • Upstream and Downstream Occlusion Protection. System occlusion above or below the pump will trigger an Air/Up Occlusion or Down Occlusion alarm.
  • Adjustable Occlusion Pressure. Configure the pump’s downstream occlusion pressure to fit your therapy.
  • Unattended Pump Alert. The pump will alert you if it has been powered on but is not infusing, and two minutes have passed without a control key being pressed.
  • Backup Power Source. When installed in the pole-mount charger, the pump switches from AC (mains) power to battery backup whenever power is interrupted.
  • Mandatory Repeat Program Verification. Whenever the infusion program ends, you must either confirm the infusion settings or change them before repeating the infusion.
  • Event Log. The pump automatically records operating events into a database that gives you an infusion history. It also helps you diagnose and prevent pump problems.
  • Latex-Free, Non-DEHP Administration Sets. For patient protection, BodyGuard® tubing sets do not contain latex or Di (2-ethylhexyl) phthalate (DEHP).
  • Dimensions
  • 4.9 x 3.5 x 1.6 in. (11.4 x 8.9 x 4.1 cm) (L/W/H)
  • Weight
  • 14.5 oz. (410 g) (with battery)
  • Infusion Modes
  • Patient-controlled epidural analgesia (PCEA) bolus-only
  • Continuous (basal) plus PCEA bolus
  • Continuous-only
  • Total Infusion Volume
  • 0.1 to 1000 mL
  • Flow Rate Range
  • 0 to 30 mL/hr (in 0.1 mL increments)
  • Delivery Accuracy
  • ±5%
  • KVO Rate
  • 0 (off) to 5 mL/hr (in 0.1 mL increments)
  • Air Sensor
  • Ultrasonic, adjustable 0.1 to 1.0 mL air bubble size (total air volume)
  • Maximum Pressure
  • Adjustable 6.8 to 21.2 psi (350 to 1100 mmHg)
  • Power Supply
  • 100-240V AC, 50/60 Hz (10W/0.3A max)
  • Battery
  • One (1) standard lithium-polymer rechargeable, 7.4V 1800 mAh
  • Battery Life
  • 10 to 12 hours at 125 mL/hr
  • Battery Charging
  • Battery charges automatically when installed in charger connected to AC (mains) power source.
  • < 8 hours required to charge fully-depleted battery.
  • Alert Conditions
  • End Program
  • Low Battery
  • End Battery
  • Pump Unattended
  • Missing Key (administration set)
  • Alarm Conditions
  • Air/Up Occlusion
  • Down Occlusion
  • Door Open
  • Error (System Malfunction)
  • BodyComm® History Download Option
  • Download and print a detailed infusion history for the pump.
  • BodyGuard® Administration Sets
  • Connect with any standard reservoir
  • Include integrated check valve
  • Non-DEHP PVC
  • Do not contain natural rubber latex